V. General
Considerations for Reprocessing Instructions
in
Device Labeling
A. Overview of
Reprocessing
Reprocessing is defined as validated processes
used to render a medical device, which has been previously used or
contaminated, fit for a subsequent single use. These
processes are designed to remove soil and contaminants by cleaning and to inactivate microorganisms by disinfection or sterilization. Reprocessing of reusable devices encompasses appropriate steps that begin in close proximity to the point of use of the device and, in general, involves the following three steps in sequence:
processes are designed to remove soil and contaminants by cleaning and to inactivate microorganisms by disinfection or sterilization. Reprocessing of reusable devices encompasses appropriate steps that begin in close proximity to the point of use of the device and, in general, involves the following three steps in sequence:
1. Point-of-Use Processing: Reprocessing
begins with processing at the point of use
(i.e., close proximity to the point of use of the device), to facilitate subsequent
(i.e., close proximity to the point of use of the device), to facilitate subsequent
cleaning steps. We define this as point-of-use processing, which
includes prompt, initial cleaning steps and/or measures to prevent drying of
soil and contaminants in and on the device.
2. Thorough Cleaning: The device should be
thoroughly cleaned after the point-of-
use processing. Generally, thorough cleaning is done in a dedicated cleaning area.
use processing. Generally, thorough cleaning is done in a dedicated cleaning area.
Devices that will likely not become
contaminated with pathogens during use (e.g., room vital signs monitor) may not
require disinfection, and therefore may be
suitable for use after cleaning only.
suitable for use after cleaning only.
3. Disinfection or Sterilization:
Depending on the intended use of the device, the
device should be disinfected or sterilized, and routed back into use.
device should be disinfected or sterilized, and routed back into use.
A simple overview of reprocessing is presented in Figure 1. A more
detailed overview of each reprocessing step is provided in Appendix B.
FIGURE 1. PROCESS OVERVIEW
Use
Use
Point-of-Use Processing
(prompt,
initial treatment to remove and/or
prevent drying of soil and contaminants)
prevent drying of soil and contaminants)
Thorough Cleaning
(and return to use, or)
Disinfection Sterilization
(Low, Intermediate,
or High Level)
or High Level)
It is important to note that cleaning, disinfection, and
sterilization are distinctly different
processes.
processes.
Cleaning is the physical removal of soil and
contaminants; the methods and agents used for cleaning should be designed to
remove such soil and contamination effectively. Effective cleaning should:
§ minimize the soil transfer from one patient
to another or between uses in a single
patient;
patient;
§ prevent
accumulation of residual soil throughout the product’s use life; and § allow for successful, subsequent
disinfection/sterilization steps.
In comparison, disinfection and sterilization
processes are intended to kill
microorganisms; the methods and agents employed
for disinfection and sterilization
should be designed to achieve appropriate
microbicidal effects. Please see
Appendix A for the definitions of disinfection and sterilization, and Section
VI. Criterion 3 for
specific information on appropriate microbicidal processes.
specific information on appropriate microbicidal processes.
Accordingly, cleaning steps should be validated separately and
independently from disinfection or sterilization steps.
An overview of reusable medical device processing is found in
Appendix B of this document.
B. Resources for
Developing Reprocessing Instructions
The following are resources to consider when developing
reprocessing instructions for reusable medical devices.
1. You should follow the labeling recommendations
in device-specific guidance,
when available. Device guidance may be found by searching FDA’s
Guidance Document Database available at
2. The following Technical Information Reports
(TIRs) developed by the
Association for the Advancement of Medical
Instrumentation (AAMI) provide technical information for manufacturers and
users and may be helpful when developing labeling instructions for reusable
medical device:
a. AAMI TIR12, Designing, testing and labeling
reusable medical devices for
reprocessing in health care settings: A guide for medical device
reprocessing in health care settings: A guide for medical device
manufacturers.
2. AAMI TIR30, A compendium of
processes, materials, test methods, and
acceptance criteria for cleaning reusable medical devices.
acceptance criteria for cleaning reusable medical devices.
3. We recommend you refer to the current
FDA-recognized version of AAMI/ANSI
ST81, Sterilization of medical devices - Information
to be provided by the
manufacturer for the processing of resterilizable medical devices.
4. We recommend you use current FDA-recognized test methods available from
manufacturer for the processing of resterilizable medical devices.
4. We recommend you use current FDA-recognized test methods available from
standards developing organizations (SDO). A
searchable database of FDA-
recognized consensus standards is available at:
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/search.cfm.
5. You should also consult any relevant clinical practice guidelines and
recommendations for infection control published by professional societies and
5. You should also consult any relevant clinical practice guidelines and
recommendations for infection control published by professional societies and
associations, standards developing
organizations, and government agencies (for
example, the “Guideline for Disinfection and
Sterilization in Healthcare Facilities,
2008” from the Centers for Disease Control (CDC), available at
http://www.cdc.gov/hicpac/pdf/guidelines/disinfection_nov_2008.pdf).
Clinical practice guidelines, however, do not always consider or correctly address
all FDA regulatory requirements. As an example, some professional organizations
may recommend using disinfectants in ways that may not necessarily comply with
FDA regulations. Compliance with FDA regulations is required.
2008” from the Centers for Disease Control (CDC), available at
http://www.cdc.gov/hicpac/pdf/guidelines/disinfection_nov_2008.pdf).
Clinical practice guidelines, however, do not always consider or correctly address
all FDA regulatory requirements. As an example, some professional organizations
may recommend using disinfectants in ways that may not necessarily comply with
FDA regulations. Compliance with FDA regulations is required.
C. Human Factors in
Developing Reprocessing Instructions
You should consider the following recommendations regarding human
factors in developing your reprocessing instructions:
1. We recommend that you develop consistent
reprocessing instructions across each
of your product lines. Labeling that provides consistent methods
and terminology, and utilizes the same document layout for all devices of a
type, may help improve the user’s comprehension and adherence to the
instructions.
2. You should address any known post-market human
factors issues known to exist
for reprocessing your device or similar
devices. Examples of human factors issues
include, but are not limited to, actions requiring substantial dexterity or strength,
good visual acuity, or familiarity with uncommon practices. Information on post-
market issues may be found by reviewing your internal user complaint files, the
published literature, the FDA’s Medical Device Reporting (MDR) system, and
FDA Safety Alerts and Public Health Notifications. We recommend that you refer
to the following sources for additional information on human factors:
include, but are not limited to, actions requiring substantial dexterity or strength,
good visual acuity, or familiarity with uncommon practices. Information on post-
market issues may be found by reviewing your internal user complaint files, the
published literature, the FDA’s Medical Device Reporting (MDR) system, and
FDA Safety Alerts and Public Health Notifications. We recommend that you refer
to the following sources for additional information on human factors:
a. FDA’s guidance “Medical Device Use-Safety:
Incorporating Human Factors
Engineering into Risk Management”
Engineering into Risk Management”
(http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidan
ce/GuidanceDocuments/ucm094461.pdf).
ce/GuidanceDocuments/ucm094461.pdf).
(http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidan
ce/GuidanceDocuments/UCM095300.pdf).
ce/GuidanceDocuments/UCM095300.pdf).
c. The current FDA-recognized version of IEC
Standard 62366, “Medical
Devices - Application of usability engineering to medical devices.”
Devices - Application of usability engineering to medical devices.”
d. The current FDA-recognized version of
ANSI/AAMI HE75, “Human Factors
Engineering - Design of Medical Devices.”
Engineering - Design of Medical Devices.”
3. For devices that are subject to design
controls under 21 CFR 820.30, you should
validate your reprocessing instructions to
ensure that users will be able to
successfully understand and follow them. FDA recommends considering the following:
a. Your validation study participants should be
representative of the professional
staff that would
perform these actual reprocessing procedures. If users would
be wearing personal protective equipment (PPE), such as goggles, full-length
face shields, heavy-duty utility gloves or liquid-resistant covering with
be wearing personal protective equipment (PPE), such as goggles, full-length
face shields, heavy-duty utility gloves or liquid-resistant covering with
sleeves, then the validation study participants
should wear them as well.
b. Participants may use the instructions to
perform an actual or simulated
reprocessing procedure or verbally describe what they would do as
they read the instructions.
c. If attributes of the use environment might
affect use of the instructions and
reprocessing of the device, they should be
represented in the study.
d. Observing and documenting participant behavior
during testing will allow you
to assess the participants’ adherence to the instructions and to
identify and understand the nature of any errors or problems that occur.
e. After using the instructions independently,
you should ask the participants if
they had difficulty in performing the
reprocessing, and allow them to describe their experience. You should ask
specifically about any errors, problems or hesitations that were observed. The
participants should provide subjective
feedback regarding any wording in the instructions that they found confusing, misleading, or incomplete. The participants’ responses and comments should be documented. If you make significant changes to the instructions after
testing them, you should validate the success of the changes at eliminating or reducing the problems previously identified.
feedback regarding any wording in the instructions that they found confusing, misleading, or incomplete. The participants’ responses and comments should be documented. If you make significant changes to the instructions after
testing them, you should validate the success of the changes at eliminating or reducing the problems previously identified.
1 comment:
As human cognition of the world increases, the level of medical care continues to increase, so is the use of medical devices. The use of medical devices is closely related to our life and health, which means that it is vital to ensure the safety and reliability of our products. Medical Device testing
Post a Comment