Principles of Bioanalytical Method Validation and Establishment
The fundamental parameters to ensure the acceptability of the performance of a
bioanalytical method validation are accuracy, precision, selectivity, sensitivity,
reproducibility, and stability.
· A specific, detailed description of the bioanalytical method should be written. This can be in
the form of a protocol, study plan, report, and/or SOP.
· Each step in the method should be investigated to determine the extent to which
environmental, matrix, material, or procedural variables can affect the estimation of analyte
in the matrix from the time of collection of the material up to and including the time of
analysis.
· It may be important to consider the variability of the matrix due to the physiological nature
of the sample. In the case of LC-MS-MS-based procedures, appropriate steps should be
taken to ensure the lack of matrix effects throughout the application of the method,
especially if the nature of the matrix changes from the matrix used during method validation.
· A bioanalytical method should be validated for the intended use or application. All
experiments used to make claims or draw conclusions about the validity of the method
should be presented in a report (method validation report).
· Whenever possible, the same biological matrix as the matrix in the intended samples should
be used for validation purposes. (For tissues of limited availability, such as bone marrow,
physiologically appropriate proxy matrices can be substituted.)
· The stability of the analyte (drug and/or metabolite) in the matrix during the collection
process and the sample storage period should be assessed, preferably prior to sample
analysis.
· For compounds with potentially labile metabolites, the stability of analyte in matrix from
dosed subjects (or species) should be confirmed.
· The accuracy, precision, reproducibility, response function, and selectivity of the method for
endogenous substances, metabolites, and known degradation products should be
established for the biological matrix. For selectivity, there should be evidence that the
substance being quantified is the intended analyte.
· The concentration range over which the analyte will be determined should be defined in the
bioanalytical method, based on evaluation of actual standard samples over the range,
including their statistical variation. This defines the standard curve.
· A sufficient number of standards should be used to adequately define the relationship
between concentration and response. The relationship between response and concentration
should be demonstrated to be continuous and reproducible. The number of standards used
should be a function of the dynamic range and nature of the concentration-response
relationship. In many cases, six to eight concentrations (excluding blank values) can define
the standard curve. More standard concentrations may be recommended for nonlinear than
for linear relationships.
· The ability to dilute samples originally above the upper limit of the standard curve should be
demonstrated by accuracy and precision parameters in the validation.
· In consideration of high throughput analyses, including but not limited to multiplexing,
multicolumn, and parallel systems, sufficient QC samples should be used to ensure control
of the assay. The number of QC samples to ensure proper control of the assay should be
determined based on the run size. The placement of QC samples should be judiciously
considered in the run.
· For a bioanalytical method to be considered valid, specific acceptance criteria should be set
in advance and achieved for accuracy and precision for the validation of QC samples over
the range of the standards.
validation refers to establishing documented evidence that a process or system, when operated within established parameters, can perform effectively and reproducibly to produce a medicinal product meeting its pre-determined specifications and quality attributes
Subscribe to:
Post Comments (Atom)
Pharmaceutical Validation Documentation Requirements
Pharmaceutical validation is a critical process that ensures that pharmaceutical products meet the desired quality standards and are safe fo...
-
K. Dashora, D. Singh, Swarnlata Saraf and S. Saraf *. Institute of Pharmacy, Pt.RavishankarShuklaUniversity, Raipur 492 010. *Author for ...
-
Validation of the Autoclave is classified into the following 1.0 OQ – Operational Qualification 2.0 PQ – Performance Qualification The valid...
-
Cold storage is a relatively simple cold room that is commonly used to store material between 2[degrees] to 8[degrees]C. Such cold rooms a...