REQUIREMENTS PRIOR TO VALIDATION
The 17665 document makes it clear in numerous locations that the user’s quality system must adhere to ISO 13485:2003 Medical devices — Quality management system — Requirements for regulatory purposes.5 So if a user wishes to claim full compliance with the new 17665 steam standard, then their quality system must also be in compliance with ISO 13485, including items such as preventive/periodic maintenance and regular calibration for the sterilizer, documentation, change control, purchasing, etc. When compared with the previous steam document, the new 17665 also has more information on product and process characterization, sterilizing agent characterization, installation qualification/IQ, and operational qualification/OQ. The new document also states more clearly that a fully compliant validation is not just a series of successful halfcycles, but is the full complement of successful IQ, OQ, and PQ.
Sterilization agent characterization will be simple for most users — moist heat/steam at 121 or 132 °C, and cycle selection (gravity, prevacuum, etc.). Process and equipment characterization means defining and documenting items like the sterilizer cycle parameters, products (or product families) to be sterilized, load configurations and limits, placement of biological indicators or chemical indicators (BIs/CIs), process tolerances, and equipment identification. Much of this type of information would be recorded in well-written validation protocols or validation final reports. Biological indicators often use spores of the bacterial species Geobacillus stearothermophilus at a titer of greater than 106per BI, although other species or titers are sometimes used.
The new 17665 document also has more information on IQ and OQ. It defines IQ as “obtaining and documenting evidence that equipment has been provided and installed in accordance with its specification.” Autoclave installations commonly document items such as the sterilizer identification numbers, location, line voltage and amperage, water supply piping and pressure limits, steam line requirements, filtration, chamber size, structure and support, piping materials, software certification, manuals, drawings and documentation, and calibrations (temperature, pressure, and timer). The sterilizer must be installed in such a manner to facilitate any necessary maintenance, repair, adjustment, cleaning, and calibration.
OQ is defined as “obtaining and documenting evidence that the installed equipment operates within predetermined limits when used in accordance with its operational procedures.” Autoclave OQs commonly test or verify items such as cycle operation and programming instructions, safety and alarm testing, error reporting, empty chamber temperature profiling and chamber temperature limits/specifications, air removal testing, leak testing, temperature control anomalies, full cycle full-load temperature profiles (if proposed fullcycle exposure time is known), and determination of any hot or cold spots within the chamber.
The product definition and process definition sections of the new document list things such as product specifications, product families, packaging, re-sterilization issues, package moisture, stability and potency of container products, re-usable container systems, process challenge devices/PCDs, sterility assurance level/SAL, BIs and CIs, and bioburden determination if necessary. PCDs are described as products or items that provide a known resistance to the sterilization process. They are commercially available or may also be created from the user’s product line by inserting spore strips, spore dots, inoculated threads, etc. into items or locations that are determined to be the most-difficult-to-sterilize product or location in the load.
There are many other activities or decisions to be made prior to or during the IQ/OQ, that are not necessarily detailed in either standard. Items such as:
- Obtaining calibrated temperature recording devices or thermocouples
- Ordering supplies such as BIs, CIs, Bowie-Dick test packs, packaging materials, etc. and noting if adequate laboratory facilities are available
- Determining worst-case validation load and worst-case test product or PCD. The protocol or final report should contain a written rationale describing how the loads and product(s) were selected
- Selecting cycle type: 121 or 132 °C, gravity or prevacuum cycle, etc.; and determining if drying time needs to be qualified
- Is product bioburden testing necessary?
- Is product resterilization to be allowed and what are the requirements for resterilization?
- Is product stability or shelf life testing necessary for the user’s products?
- Does packaging testing or packaging validation need to be included with the protocol?
VALIDATION – PERFORMANCE QUALIFICATION
AAMI TIR #13 states “Sterilization process validation is a documented procedure for obtaining, recording, and interpreting the results required to establish that a process will consistently yield product complying with its predetermined specifications.” For the purposes of this article, the primary specification will be sterility. The performance qualification/PQ or microbiological qualification is a series of tests that establishes that the installed and properly operating sterilizer will process the users desired chamber loads to achieve the specified sterility assurance level/SAL. It must be remembered that the load is part of the validation — that is, if the user makes significant changes to the load at any point in the future — then re-validation may be necessary. The previous ISO 11134 document gave relatively little guidance information and few specifications for conducting the test cycles necessary to qualify the user’s proposed fullcycle exposure time(s). The new 17665 steam document varies little from the previous standard in respect to the minimal PQ information that is provided. The 17665 describes bioburden validation methods and the more commonly used halfcycle “overkill” method. It should be noted that at the time this article was prepared, the proposed guidance document that is to accompany ISO 17665-1 was not yet available. This guidance document may provide more advice on microbiological qualification issues (ISO 17665-2 Sterilization of health care products — Moist heat — Part 2: Guidance on the application of ISO 17665-1). For this article, the general requirements for an overkill cycle PQ will be reviewed.
While many activities are required to complete the PQ, the primary goal for the commonly employed overkill validation is this: the user needs to complete three consecutive successful halfcycles in order to qualify their proposed fullcycle exposure for routine processing of sterilization loads. In our case, successful means all BIs are killed (no growth upon incubation) for the three consecutive halfcycles. If, for example, there was no BI growth for the three test cycles at ten minutes exposure at 121 °C, then a 20-minute exposure at the same temperature would be adequate for routine daily processing, assuming all other aspects or requirements of the IQ/OQ/PQ are successful, documented, reviewed, and approved.
But a description of the PQ needs much more detail than this. Validation protocols vary in format from company to company, but most will capture similar information for the final report. An example of validation protocol and final report sections would be:
- Title page with approval signatures
- Purpose, background information, or general goal(s) of validation
- Scope with more specifics about methods, cycles, facility, SAL, products and load, exclusions, etc.
- References with published standards and company SOPs
- Equipment, supplies, validation loads, BIs, etc.
- Rationale for selection of products, load, cycles, PCDs, etc.
- Procedure or methods (more details on this below)
- Acceptance criteria which list the pass/fail requirements
- Deviation report which lists any unexpected results, with potential effects on the validation, along with accept/reject rationale
- Results and conclusions which assign a pass/fail decision to each acceptance criteria, summarize study, and include any requirements for revalidation
- Attachment which lists any data sheets, diagrams, certificates, temperature records, etc., for inclusion with final report
- Approvals section for final report.
Final reports should contain: 1) all sterilizer run data or recorder charts, signed and reviewed; 2) all temperature recorder data, signed and reviewed; 3) all data sheets with BI, CI, or any other test results, reviewed and signed; 4) any deviations recorded and investigated, with final disposition; 5) results, conclusions, and discussion; 6) calibration documents for any measuring instruments used during the study; 7) the approved full-cycle parameters and acceptable placement locations for BIs for normal processing; and 8) manufacturers’ certificates of analysis for any items such as BIs, growth media, growth promotion test cultures, etc. Including digital photographs of sterilizer, load, PCD, etc. can be quite helpful for an auditor who may be reviewing the report at a later date. The completed final report packet must then be routed for review and signed for approval.
POST-VALIDATION
There are still issues to be addressed when all activities seem to have been completed. The sterilizer must be added to a regular and documented calibration program. The sterilizer must be included in a regular and documented periodic/preventive maintenance program. And the sterilizer must be added to the validation schedule for its annual requalification. The user needs to verify that all personnel that will be using the autoclave are trained using applicable operation and safety SOPs. Untrained staff should not be allowed to run the sterilizer. Approved products, loads, cycles, and load limit information must be readily available to all operators. SOPs for daily processing must list all requirements for data that is to be reviewed and retained from the sterilizer runs, with logbook, filing system, or archive for run records. SOPs must also address items such as 1) segregation of processed and non-processed product, 2) storage requirements for processed products if necessary, 3) notification of management or maintenance if sterilizer malfunctions or if recorder chart lists any errors, cautions, or warnings, 4) immediate notification of management for BI test failure, including investigation and product quarantine procedure as appropriate, and 5) resterilization requirements if resteril-ization is to be allowed.
In summary, there seem to be no drastic or revolutionary changes in making the transition from ISO 11134 to ISO 17665. The new 17665 steam document provides more information and more guidance in some areas, while leaving other areas (such as PQ) relatively unchanged. While users would be advised to obtain the 17665-2 guidance document when it becomes available, it is anticipated that manufacturers will not find any great difficulties in applying the new standard.
References
- ISO 11134:1994. Sterilization of health care products — Requirements for validation and routine control — Industrial moist heat sterilization.
- ANSI/AAMI/ISO 17665-1:2006. Sterilization of health care products — Moist heat — Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
- AAMI TIR No. 13-1997. Principles of Industrial Moist Heat Sterilization.
- PDA Technical Report #1. Validation of Steam Sterilization Cycles. Parenteral Drug Association.
- ISO 13485:2003. Medical devices — Quality management systems — Requirements for regulatory purposes.
3 comments:
Compared to other sterilization methods like CIP or SIP, steam sterilizer provides an economical solution to sterilize medical waste, infectious waste and other materials. This machine not only helps to eliminate the problem of environmental pollution by doing away with methods like incineration but also helps to minimize and eliminate the expense of disposing off untreated medical waste.Stem sterilizers carry so many advantages, but has many disadvantages also.Find more about steam sterilizers.
I always waiting for your post please update it thanx Autoclaves Products Suppliers
Thanks for sharing such a great blog Keep posting..
chứng nhận iso 13485
Post a Comment