Tuesday, December 1, 2009

Bioanalytical Method Validation

Table of Contents
I. INTRODUCTION.................................................................................................................................................................1
II. BACKGROUND...................................................................................................................................................................1
A. FULL VALIDATION .............................................................................................................................................................2
B. PARTIAL VALIDATION .......................................................................................................................................................2
C. CROSS-VALIDATION ..........................................................................................................................................................3
III. REFERENCE STANDARD............................................................................................................................................4
IV. METHOD DEVELOPMENT: CHEMICAL ASSAY........................................................................................................4
A. SELECTIVITY......................................................................................................................................................................4
B. ACCURACY, PRECISION, AND RECOVERY........................................................................................................................5
C. CALIBRATION/STANDARD CURVE.....................................................................................................................................5
D. STABILITY..........................................................................................................................................................................6
E. PRINCIPLES OF BIOANALYTICAL METHOD VALIDATION AND ESTABLISHMENT ............................................................8
F. SPECIFIC RECOMMENDATIONS FOR METHOD VALIDATION........................................................................................... 10
V. METHOD DEVELOPMENT: MICROBIOLOGICAL AND LIGAND-BINDING ASSAYS ..................................11
A. SELECTIVITY ISSUES ....................................................................................................................................................... 11
B. QUANTIFICATION ISSUES ................................................................................................................................................ 12
VI. APPLICATION OF VALIDATED METHOD TO ROUTINE DRUG ANALYSIS....................................................13
ACCEPTANCE CRITERIA FOR THE RUN..................................................................................................................................... 15
VII. DOCUMENTATION.....................................................................................................................................................16
A. SUMMARY INFORMATION................................................................................................................................................ 16
B. DOCUMENTATION FOR METHOD ESTABLISHMENT ....................................................................................................... 17
C. APPLICATION TO ROUTINE DRUG ANALYSIS ................................................................................................................. 17
D. OTHER INFORMATION..................................................................................................................................................... 19
GLOSSARY...................................................................................................................................................................................20
GUIDANCE FOR INDUSTRY1
Bioanalytical Method Validation
I. INTRODUCTION
This guidance provides assistance to sponsors of investigational new drug applications (INDs), new
drug applications (NDAs), abbreviated new drug applications (ANDAs), and supplements in
developing bioanalytical method validation information used in human clinical pharmacology,
bioavailability (BA), and bioequivalence (BE) studies requiring pharmacokinetic (PK) evaluation. This
guidance also applies to bioanalytical methods used for non-human pharmacology/toxicology studies
and preclinical studies. For studies related to the veterinary drug approval process, this guidance
applies only to blood and urine BA, BE, and PK studies.
The information in this guidance generally applies to bioanalytical procedures such as gas
chromatography (GC), high-pressure liquid chromatography (LC), combined GC and LC mass
spectrometric (MS) procedures such as LC-MS, LC-MS-MS, GC-MS, and GC-MS-MS performed
for the quantitative determination of drugs and/or metabolites in biological matrices
such as blood, serum, plasma, or urine. This guidance also applies to other bioanalytical methods, such
as immunological and microbiological procedures, and to other biological matrices, such as tissue and
skin samples.
This guidance provides general recommendations for bioanalytical method validation. The
recommendations can be adjusted or modified depending on the specific type of analytical method used.
II. BACKGROUND
1 This guidance has been prepared by the Biopharmaceutics Coordinating Committee in the Center for Drug Evaluation and
Research (CDER) in cooperation with the Center for Veterinary Medicine (CVM) at the Food and Drug Administration.
This guidance represents the Food and Drug Administration's current thinking on this topic. It
does not create or confer any rights for or on any person and does not operate to bind FDA or the
public. An alternative approach may be used if such approach satisfies the requirements of the
applicable statutes and regulations.
2
This guidance has been developed based on the deliberations of two workshops: (1) Analytical
Methods Validation: Bioavailability, Bioequivalence, and Pharmacokinetic Studies (held on December
3B5, 19902 ) and (2) Bioanalytical Methods Validation C A Revisit With a Decade of Progress (held
on January 12B14, 20003).
Selective and sensitive analytical methods for the quantitative evaluation of drugs and their metabolites
(analytes) are critical for the successful conduct of preclinical and/or biopharmaceutics and clinical
pharmacology studies. Bioanalytical method validation includes all of the procedures that demonstrate
that a particular method used for quantitative measurement of analytes in a given biological matrix, such
as blood, plasma, serum, or urine, is reliable and reproducible for the intended use. The fundamental
parameters for this validation include (1) accuracy, (2) precision, (3) selectivity, (4) sensitivity, (5)
reproducibility, and (6) stability. Validation involves documenting, through the use of specific laboratory
investigations, that the performance characteristics of the method are suitable and reliable for the
intended analytical applications. The acceptability of analytical data corresponds directly to the criteria
used to validate the method.
Published methods of analysis are often modified to suit the requirements of the laboratory performing
the assay. These modifications should be validated to ensure suitable performance of the analytical
method. When changes are made to a previously validated method, the analyst should exercise
judgment as to how much additional validation is needed. During the course of a typical drug
development program, a defined bioanalytical method undergoes many modifications. The evolutionary
changes to support specific studies and different levels of validation demonstrate the validity of an
assay’s performance. Different types and levels of validation are defined and characterized as follows:
A. Full Validation
· Full validation is important when developing and implementing a bioanalytical method for
the first time.
· Full validation is important for a new drug entity.
· A full validation of the revised assay is important if metabolites are added to an existing
assay for quantification.
B. Partial Validation
Partial validations are modifications of already validated bioanalytical methods. Partial validation
can range from as little as one intra-assay accuracy and precision determination to a nearly full
2 Workshop Report: Shah, V.P. et al., Pharmaceutical Research: 1992; 9:588-592.
3 Workshop Report: Shah, V.P. et al., Pharmaceutical Research: 2000; 17:in press.
3
validation. Typical bioanalytical method changes that fall into this category include, but are not
limited to:
· Bioanalytical method transfers between laboratories or analysts
· Change in analytical methodology (e.g., change in detection systems)
· Change in anticoagulant in harvesting biological fluid
· Change in matrix within species (e.g., human plasma to human urine)
· Change in sample processing procedures
· Change in species within matrix (e.g., rat plasma to mouse plasma)
· Change in relevant concentration range
· Changes in instruments and/or software platforms
· Limited sample volume (e.g., pediatric study)
· Rare matrices
· Selectivity demonstration of an analyte in the presence of concomitant medications
· Selectivity demonstration of an analyte in the presence of specific metabolites
C. Cross-Validation
Cross-validation is a comparison of validation parameters when two or more bioanalytical methods
are used to generate data within the same study or across different studies. An example of crossvalidation
would be a situation where an original validated bioanalytical method serves as the
reference and the revised bioanalytical method is the comparator. The comparisons should be
done both ways.
When sample analyses within a single study are conducted at more than one site or more than one
laboratory, cross-validation with spiked matrix standards and subject samples should be conducted
at each site or laboratory to establish interlaboratory reliability. Cross-validation should also be
considered when data generated using different analytical techniques (e.g., LC-MS-MS vs.
ELISA4) in different studies are included in a regulatory submission.
All modifications should be assessed to determine the recommended degree of validation. The
analytical laboratory conducting pharmacology/toxicology and other preclinical studies for regulatory
submissions should adhere to FDA=s Good Laboratory Practices (GLPs)5 (21 CFR part 58) and to
sound principles of quality assurance throughout the testing process. The bioanalytical method for
human BA, BE, PK, and drug interaction studies must meet the criteria in 21 CFR 320.29. The
analytical laboratory should have a written set of standard operating procedures (SOPs) to ensure a
complete system of quality control and assurance. The SOPs should cover all aspects of analysis
from the time the sample is collected and reaches the laboratory until the results of the analysis are
reported. The SOPs also should include record keeping, security and chain of sample custody
4 Enzyme linked immune sorbent assay
5 For the Center for Veterinary Medicine, all bioequivalence studies are subject to Good Laboratory Practices.
4
(accountability systems that ensure integrity of test articles), sample preparation, and analytical tools
such as methods, reagents, equipment, instrumentation, and procedures for quality control and
verification of results.
The process by which a specific bioanalytical method is developed, validated, and used in routine
sample analysis can be divided into (1) reference standard preparation, (2) bioanalytical method
development and establishment of assay procedure, and (3) application of validated bioanalytical
method to routine drug analysis and acceptance criteria for the analytical run and/or batch. These
three processes are described in the following sections of this guidance.
III. REFERENCE STANDARD
Analysis of drugs and their metabolites in a biological matrix is carried out using samples spiked with
calibration (reference) standards and using quality control (QC) samples. The purity of the reference
standard used to prepare spiked samples can affect study data. For this reason, an authenticated
analytical reference standard of known identity and purity should be used to prepare solutions of known
concentrations. If possible, the reference standard should be identical to the analyte. When this is not
possible, an established chemical form (free base or acid, salt or ester) of known purity can be used.
Three types of reference standards are usually used: (1) certified reference standards (e.g., USP
compendial standards); (2) commercially supplied reference standards obtained from a reputable
commercial source; and/or (3) other materials of documented purity custom-synthesized by an analytical
laboratory or other noncommercial establishment. The source and lot number, expiration date,
certificates of analyses when available, and/or internally or externally generated evidence of identity and
purity should be furnished for each reference standard.
IV. METHOD DEVELOPMENT: CHEMICAL ASSAY
The method development and establishment phase defines the chemical assay. The fundamental
parameters for a bioanalytical method validation are accuracy, precision, selectivity, sensitivity,
reproducibility, and stability. Measurements for each analyte in the biological matrix should be
validated. In addition, the stability of the analyte in spiked samples should be determined. Typical
method development and establishment for a bioanalytical method include determination of (1)
selectivity, (2) accuracy, precision, recovery, (3) calibration curve, and (4) stability of analyte in spiked
samples.
A. Selectivity
Selectivity is the ability of an analytical method to differentiate and quantify the analyte in the
presence of other components in the sample. For selectivity, analyses of blank samples of the
appropriate biological matrix (plasma, urine, or other matrix) should be obtained from at least
5
six sources. Each blank sample should be tested for interference, and selectivity should be
ensured at the lower limit of quantification (LLOQ).
Potential interfering substances in a biological matrix include endogenous matrix components,
metabolites, decomposition products, and in the actual study, concomitant medication and other
exogenous xenobiotics. If the method is intended to quantify more than one analyte, each
analyte should be tested to ensure that there is no interference.
B. Accuracy, Precision, and Recovery
The accuracy of an analytical method describes the closeness of mean test results obtained by
the method to the true value (concentration) of the analyte. Accuracy is determined by replicate
analysis of samples containing known amounts of the analyte. Accuracy should be measured
using a minimum of five determinations per concentration. A minimum of three concentrations in
the range of expected concentrations is recommended. The mean value should be within 15%
of the actual value except at LLOQ, where it should not deviate by more than 20%. The
deviation of the mean from the true value serves as the measure of accuracy.
The precision of an analytical method describes the closeness of individual measures of an
analyte when the procedure is applied repeatedly to multiple aliquots of a single homogeneous
volume of biological matrix. Precision should be measured using a minimum of five
determinations per concentration. A minimum of three concentrations in the range of expected
concentrations is recommended. The precision determined at each concentration level should
not exceed 15% of the coefficient of variation (CV) except for the LLOQ, where it should not
exceed 20% of the CV. Precision is further subdivided into within-run, intra-batch precision or
repeatability, which assesses precision during a single analytical run, and between-run, interbatch
precision or repeatability, which measures precision with time, and may involve different
analysts, equipment, reagents, and laboratories.
The recovery of an analyte in an assay is the detector response obtained from an amount of the
analyte added to and extracted from the biological matrix, compared to the detector response
obtained for the true concentration of the pure authentic standard. Recovery pertains to the
extraction efficiency of an analytical method within the limits of variability. Recovery of the
analyte need not be 100%, but the extent of recovery of an analyte and of the internal standard
should be consistent, precise, and reproducible. Recovery experiments should be performed by
comparing the analytical results for extracted samples at three concentrations (low, medium, and
high) with unextracted standards that represent 100% recovery.
C. Calibration/Standard Curve
A calibration (standard) curve is the relationship between instrument response and known
concentrations of the analyte. A calibration curve should be generated for each analyte in the
6
sample. A sufficient number of standards should be used to adequately define the relationship
between concentration and response. A calibration curve should be prepared in the same
biological matrix as the samples in the intended study by spiking the matrix with known
concentrations of the analyte. The number of standards used in constructing a calibration curve
will be a function of the anticipated range of analytical values and the nature of the
analyte/response relationship. Concentrations of standards should be chosen on the basis of the
concentration range expected in a particular study. A calibration curve should consist of a blank
sample (matrix sample processed without internal standard), a zero sample (matrix sample
processed with internal standard), and six to eight non-zero samples covering the expected
range, including LLOQ.
1. Lower Limit of Quantification (LLOQ)
The lowest standard on the calibration curve should be accepted as the limit of
quantification if the following conditions are met:
C The analyte response at the LLOQ should be at least 5 times the response
compared to blank response.
C Analyte peak (response) should be identifiable, discrete, and reproducible with
a precision of 20% and accuracy of 80-120%.
2. Calibration Curve/Standard Curve/Concentration-Response
The simplest model that adequately describes the concentration-response relationship
should be used. Selection of weighting and use of a complex regression equation should
be justified. The following conditions should be met in developing a calibration curve:
C #20% deviation of the LLOQ from nominal concentration
C #15% deviation of standards other than LLOQ from nominal concentration
At least four out of six non-zero standards should meet the above criteria, including the
LLOQ and the calibration standard at the highest concentration. Excluding the
standards should not change the model used.
D. Stability
Drug stability in a biological fluid is a function of the storage conditions, the chemical properties
of the drug, the matrix, and the container system. The stability of an analyte in a particular
matrix and container system is relevant only to that matrix and container system and should not
be extrapolated to other matrices and container systems. Stability procedures should evaluate
the stability of the analytes during sample collection and handling, after long-term (frozen at the
7
intended storage temperature) and short-term (bench top, room temperature) storage, and after
going through freeze and thaw cycles and the analytical process. Conditions used in stability
experiments should reflect situations likely to be encountered during actual sample handling and
analysis. The procedure should also include an evaluation of analyte stability in stock solution.
All stability determinations should use a set of samples prepared from a freshly made stock
solution of the analyte in the appropriate analyte-free, interference-free biological matrix. Stock
solutions of the analyte for stability evaluation should be prepared in an appropriate solvent at
known concentrations.
1. Freeze and Thaw Stability
Analyte stability should be determined after three freeze and thaw cycles. At least three
aliquots at each of the low and high concentrations should be stored at the intended
storage temperature for 24 hours and thawed unassisted at room temperature. When
completely thawed, the samples should be refrozen for 12 to 24 hours under the same
conditions. The freeze–thaw cycle should be repeated two more times, then analyzed
on the third cycle. If an analyte is unstable at the intended storage temperature, the
stability sample should be frozen at -700C during the three freeze and thaw cycles.
2. Short-Term Temperature Stability
Three aliquots of each of the low and high concentrations should be thawed at room
temperature and kept at this temperature from 4 to 24 hours (based on the expected
duration that samples will be maintained at room temperature in the intended study) and
analyzed.
3. Long-Term Stability
The storage time in a long-term stability evaluation should exceed the time between the
date of first sample collection and the date of last sample analysis. Long-term stability
should be determined by storing at least three aliquots of each of the low and high
concentrations under the same conditions as the study samples. The volume of samples
should be sufficient for analysis on three separate occasions. The concentrations of all
the stability samples should be compared to the mean of back-calculated values for the
standards at the appropriate concentrations from the first day of long-term stability
testing.
4. Stock Solution Stability
The stability of stock solutions of drug and the internal standard should be evaluated at
room temperature for at least 6 hours. If the stock solutions are refrigerated or frozen
8
for the relevant period, the stability should be documented. After completion of the
desired storage time, the stability should be tested by comparing the instrument
response with that of freshly prepared solutions.
5. Post-Preparative Stability
The stability of processed samples, including the resident time in the autosampler, should
be determined. The stability of the drug and the internal standard should be assessed
over the anticipated run time for the batch size in validation samples by determining
concentrations on the basis of original calibration standards.
Although the traditional approach of comparing analytical results for stored samples with
those for freshly prepared samples has been referred to in this guidance, other statistical
approaches based on confidence limits for evaluation of an analyte=s stability in a
biological matrix can be used. SOPs should clearly describe the statistical method and
rules used. Additional validation may include investigation of samples from dosed
subjects.
E. Principles of Bioanalytical Method Validation and Establishment
· The fundamental parameters to ensure the acceptability of the performance of a
bioanalytical method validation are accuracy, precision, selectivity, sensitivity,
reproducibility, and stability.
· A specific, detailed description of the bioanalytical method should be written. This can be in
the form of a protocol, study plan, report, and/or SOP.
· Each step in the method should be investigated to determine the extent to which
environmental, matrix, material, or procedural variables can affect the estimation of analyte
in the matrix from the time of collection of the material up to and including the time of
analysis.
· It may be important to consider the variability of the matrix due to the physiological nature
of the sample. In the case of LC-MS-MS-based procedures, appropriate steps should be
taken to ensure the lack of matrix effects throughout the application of the method,
especially if the nature of the matrix changes from the matrix used during method validation.
· A bioanalytical method should be validated for the intended use or application. All
experiments used to make claims or draw conclusions about the validity of the method
should be presented in a report (method validation report).
9
· Whenever possible, the same biological matrix as the matrix in the intended samples should
be used for validation purposes. (For tissues of limited availability, such as bone marrow,
physiologically appropriate proxy matrices can be substituted.)
· The stability of the analyte (drug and/or metabolite) in the matrix during the collection
process and the sample storage period should be assessed, preferably prior to sample
analysis.
· For compounds with potentially labile metabolites, the stability of analyte in matrix from
dosed subjects (or species) should be confirmed.
· The accuracy, precision, reproducibility, response function, and selectivity of the method for
endogenous substances, metabolites, and known degradation products should be
established for the biological matrix. For selectivity, there should be evidence that the
substance being quantified is the intended analyte.
· The concentration range over which the analyte will be determined should be defined in the
bioanalytical method, based on evaluation of actual standard samples over the range,
including their statistical variation. This defines the standard curve.
· A sufficient number of standards should be used to adequately define the relationship
between concentration and response. The relationship between response and concentration
should be demonstrated to be continuous and reproducible. The number of standards used
should be a function of the dynamic range and nature of the concentration-response
relationship. In many cases, six to eight concentrations (excluding blank values) can define
the standard curve. More standard concentrations may be recommended for nonlinear than
for linear relationships.
· The ability to dilute samples originally above the upper limit of the standard curve should be
demonstrated by accuracy and precision parameters in the validation.
· In consideration of high throughput analyses, including but not limited to multiplexing,
multicolumn, and parallel systems, sufficient QC samples should be used to ensure control
of the assay. The number of QC samples to ensure proper control of the assay should be
determined based on the run size. The placement of QC samples should be judiciously
considered in the run.
· For a bioanalytical method to be considered valid, specific acceptance criteria should be set
in advance and achieved for accuracy and precision for the validation of QC samples over
the range of the standards.
10
F. Specific Recommendations for Method Validation
· The matrix-based standard curve should consist of a minimum of six standard points,
excluding blanks, using single or replicate samples. The standard curve should cover the
entire range of expected concentrations.
· Standard curve fitting is determined by applying the simplest model that adequately
describes the concentration-response relationship using appropriate weighting and statistical
tests for goodness of fit.
· LLOQ is the lowest concentration of the standard curve that can be measured with
acceptable accuracy and precision. The LLOQ should be established using at least five
samples independent of standards and determining the coefficient of variation and/or
appropriate confidence interval. The LLOQ should serve as the lowest concentration on
the standard curve and should not be confused with the limit of detection and/or the low QC
sample. The highest standard will define the upper limit of quantification (ULOQ) of an
analytical method.
· For validation of the bioanalytical method, accuracy and precision should be determined
using a minimum of five determinations per concentration level (excluding blank samples).
The mean value should be within ±15% of the theoretical value, except at LLOQ, where it
should not deviate by more than ±20%. The precision around the mean value should not
exceed 15% of the CV, except for LLOQ, where it should not exceed 20% of the CV.
Other methods of assessing accuracy and precision that meet these limits may be equally
acceptable.
· The accuracy and precision with which known concentrations of analyte in biological matrix
can be determined should be demonstrated. This can be accomplished by analysis of
replicate sets of analyte samples of known concentrations C QC samples C from an
equivalent biological matrix. At a minimum, three concentrations representing the entire
range of the standard curve should be studied: one within 3x the lower limit of quantification
(LLOQ) (low QC sample), one near the center (middle QC), and one near the upper
boundary of the standard curve (high QC).
· Reported method validation data and the determination of accuracy and precision should
include all outliers; however, calculations of accuracy and precision excluding values that are
statistically determined as outliers can also be reported.
· The stability of the analyte in biological matrix at intended storage temperatures should be
established. The influence of freeze-thaw cycles (a minimum of three cycles at two
concentrations in triplicate) should be studied.
11
· The stability of the analyte in matrix at ambient temperature should be evaluated over a time
period equal to the typical sample preparation, sample handling, and analytical run times.
· Reinjection reproducibility should be evaluated to determine if an analytical run could be
reanalyzed in the case of instrument failure.
· The specificity of the assay methodology should be established using a minimum of six
independent sources of the same matrix. For hyphenated mass spectrometry-based
methods, however, testing six independent matrices for interference may not be important.
In the case of LC-MS and LC-MS-MS-based procedures, matrix effects should be
investigated to ensure that precision, selectivity, and sensitivity will not be compromised.
Method selectivity should be evaluated during method development and throughout method
validation and can continue throughout application of the method to actual study samples.
· Acceptance/rejection criteria for spiked, matrix-based calibration standards and validation
QC samples should be based on the nominal (theoretical) concentration of analytes.
Specific criteria can be set up in advance and achieved for accuracy and precision over the
range of the standards, if so desired.
V. METHOD DEVELOPMENT: MICROBIOLOGICAL AND LIGAND-BINDING
ASSAYS
Many of the bioanalytical validation parameters and principles discussed above are also applicable to
microbiological and ligand-binding assays. However, these assays possess some unique characteristics
that should be considered during method validation.
A. Selectivity Issues
As with chromatographic methods, microbiological and ligand-binding assays should be shown
to be selective for the analyte. The following recommendations for dealing with two selectivity
issues should be considered:
1. Interference From Substances Physiochemically Similar to the Analyte
· Cross-reactivity of metabolites, concomitant medications, or endogenous
compounds should be evaluated individually and in combination with the analyte
of interest.
· When possible, the immunoassay should be compared with a validated reference
method (such as LC-MS) using incurred samples and predetermined criteria for
agreement of accuracy of immunoassay and reference method.
12
· The dilutional linearity to the reference standard should be assessed using study
(incurred) samples.
· Selectivity may be improved for some analytes by incorporation of separation
steps prior to immunoassay.
2. Matrix effects Unrelated to the Analyte
· The standard curve in biological fluids should be compared with standard in
buffer to detect matrix effects.
· Parallelism of diluted study samples should be evaluated with diluted standards to
detect matrix effects.
· Nonspecific binding should be determined.
B. Quantification Issues
Microbiological and imunoassay standard curves are inherently nonlinear and, in general, more
concentration points may be recommended to define the fit over the standard curve range than
for chemical assays. In addition to their nonlinear characteristics, the response-error
relationship for immunoassay standard curves is a nonconstant function of the mean response
(heteroscadisticity). For these reasons, a minimum of six non-zero calibrator concentrations,
run in duplicate, is recommended. The concentration-response relationship is most often fitted
to a 4- or 5-parameter logistic model, although others may be used with suitable validation. The
use of anchoring points in the asymptotic high- and low-concentration ends of the standard
curve may improve the overall curve fit. Generally, these anchoring points will be at
concentrations that are below the established LLOQ and above the established ULOQ.
Whenever possible, calibrators should be prepared in the same matrix as the study samples or
in an alternate matrix of equivalent performance. Both ULOQ and LLOQ should be defined by
acceptable accuracy, precision, or confidence interval criteria based on the study requirements.
For all assays the key factor is the accuracy of the reported results. This accuracy can be
improved by the use of replicate samples. In the case where replicate samples should be
measured during the validation to improve accuracy, the same procedure should be followed as
for unknown samples.
The following recommendations apply to quantification issues:
· If separation is used prior to assay for study samples but not for standards, it is important to
establish recovery and use it in determining results. Possible approaches to assess efficiency
13
and reproducibility of recovery are (1) the use of radiolabeled tracer analyte (quantity too
small to affect the assay), (2) the advance establishment of reproducible recovery, (3) the
use of an internal standard that is not recognized by the antibody but can be measured by
another technique.
· Key reagents, such as antibody, tracer, reference standard, and matrix should be
characterized appropriately and stored under defined conditions.
· Assessments of analyte stability should be conducted in true study matrix (e.g., should not
use a matrix stripped to remove endogenous interferences).
· Acceptance criteria: At least 67% (4 out of 6) of QC samples should be within 15% of
their respective nominal value, 33% of the QC samples (not all replicates at the same
concentration) may be outside 15% of nominal value. In certain situations, wider
acceptance criteria may be justified.
· Assay reoptimization or validation may be important when there are changes in key
reagents, as follows:
Labeled analyte (tracer)
· Binding should be reoptimized.
· Performance should be verified with standard curve and QCs.
Antibody
· Key cross-reactivities should be checked.
· Tracer experiments above should be repeated.
Matrix
· Tracer experiments above should be repeated.
Method development experiments should include a minimum of six runs conducted over several days,
with at least four concentrations (LLOQ, low, medium, and high) analyzed in duplicate in each run.
VI. APPLICATION OF VALIDATED METHOD TO ROUTINE DRUG ANALYSIS
Assays of all samples of an analyte in a biological matrix should be completed within the time period for
which stability data are available. In general, biological samples can be analyzed with a single
determination without duplicate or replicate analysis if the assay method has acceptable variability as
defined by validation data. This is true for procedures where precision and accuracy variabilities
routinely fall within acceptable tolerance limits. For a difficult procedure with a labile analyte where high
precision and accuracy specifications may be difficult to achieve, duplicate or even triplicate analyses
can be performed for a better estimate of analyte.
14
A calibration curve should be generated for each analyte to assay samples in each analytical run and
should be used to calculate the concentration of the analyte in the unknown samples in the run. The
spiked samples can contain more than one analyte. An analytical run can consist of QC samples,
calibration standards, and either (1) all the processed samples to be analyzed as one batch or (2) a
batch composed of processed unknown samples of one or more volunteers in a study. The calibration
(standard) curve should cover the expected unknown sample concentration range in addition to a
calibrator sample at LLOQ. Estimation of concentration in unknown samples by extrapolation of
standard curves below LLOQ or above the highest standard is not recommended. Instead, the
standard curve should be redefined or samples with higher concentration should be diluted and
reassayed. It is preferable to analyze all study samples from a subject in a single run.
Once the analytical method has been validated for routine use, its accuracy and precision should be
monitored regularly to ensure that the method continues to perform satisfactorily. To achieve this
objective, a number of QC samples prepared separately should be analyzed with processed test
samples at intervals based on the total number of samples. The QC samples in duplicate at three
concentrations (one near the LLOQ (i.e., #3 x LLOQ), one in midrange, and one close to the high end
of the range) should be incorporated in each assay run. The number of QC samples (in multiples of
three) will depend on the total number of samples in the run. The results of the QC samples provide the
basis of accepting or rejecting the run. At least four of every six QC samples should be within "15% of
their respective nominal value. Two of the six QC samples may be outside the "15% of their respective
nominal value, but not both at the same concentration.
The following recommendations should be noted in applying a bioanalytical method to routine drug
analysis:
· A matrix-based standard curve should consist of a minimum of six standard points,
excluding blanks (either single or replicate), covering the entire range.
· Response Function: Typically, the same curve fitting, weighting, and goodness of fit
determined during prestudy validation should be used for the standard curve within the
study. Response function is determined by appropriate statistical tests based on the actual
standard points during each run in the validation. Changes in the response function
relationship between prestudy validation and routine run validation indicate potential
problems.
· The QC samples should be used to accept or reject the run. These QC samples are matrix
spiked with analyte.
· System suitability: Based on the analyte and technique, a specific SOP (or sample) should
be identified to ensure optimum operation of the system used.
15
· Any required sample dilutions should use like matrix (e.g., human to human) obviating the
need to incorporate actual within-study dilution matrix QC samples.
· Repeat Analysis: It is important to establish an SOP or guideline for repeat analysis and
acceptance criteria. This SOP or guideline should explain the reasons for repeating sample
analysis. Reasons for repeat analyses could include repeat analysis of clinical or preclinical
samples for regulatory purposes, inconsistent replicate analysis, samples outside of the assay
range, sample processing errors, equipment failure, poor chromatography, and inconsistent
pharmacokinetic data. Reassays should be done in triplicate if sample volume allows. The
rationale for the repeat analysis and the reporting of the repeat analysis should be clearly
documented.
· Sample Data Reintegration: An SOP or guideline for sample data reintegration should be
established. This SOP or guideline should explain the reasons for reintegration and how the
reintegration is to be performed. The rationale for the reintegration should be clearly
described and documented. Original and reintegration data should be reported.
Acceptance Criteria for the Run
The following acceptance criteria should be considered for accepting the analytical run:
· Standards and QC samples can be prepared from the same spiking stock solution,
provided the solution stability and accuracy have been verified. A single source of matrix
may also be used, provided selectivity has been verified.
· Standard curve samples, blanks, QCs, and study samples can be arranged as considered
appropriate within the run.
· Placement of standards and QC samples within a run should be designed to detect assay
drift over the run.
· Matrix-based standard calibration samples: 75%, or a minimum of six standards, when
back-calculated (including ULOQ) should fall within ±15%, except for LLOQ, when it
should be ±20% of the nominal value. Values falling outside these limits can be discarded,
provided they do not change the established model.
· Acceptance criteria for accuracy and precision as outlined in section IV.F, “Specific
Recommendation for Method Validation,” should be provided for both the intra-day and
intra-run experiment.
· Quality Control Samples: Quality control samples replicated (at least once) at a minimum of
three concentrations (one within 3x of the LLOQ (low QC), one in the midrange (middle
16
QC), and one approaching the high end of the range (high QC)) should be incorporated into
each run. The results of the QC samples provide the basis of accepting or rejecting the run.
At least 67% (four out of six) of the QC samples should be within 15% of their respective
nominal (theoretical) values; 33% of the QC samples (not all replicates at the same
concentration) can be outside the ±15% of the nominal value. A confidence interval
approach yielding comparable accuracy and precision is an appropriate alternative.
The minimum number of samples (in multiples of three) should be at least 5% of the number of
unknown samples or six total QCs, whichever is greater.
· Samples involving multiple analytes should not be rejected based on the data from one
analyte failing the acceptance criteria.
· The data from rejected runs need not be documented, but the fact that a run was rejected
and the reason for failure should be recorded.
VII. DOCUMENTATION
The validity of an analytical method should be established and verified by laboratory studies, and
documentation of successful completion of such studies should be provided in the assay validation
report. General and specific SOPs and good record keeping are an essential part of a validated
analytical method. The data generated for bioanalytical method establishment and the QCs should be
documented and available for data audit and inspection. Documentation for submission to the Agency
should include (1) summary information, (2) method development and establishment, (3) bioanalytical
reports of the application of any methods to routine sample analysis, and (4) other information
applicable to method development and establishment and/or to routine sample analysis.
A. Summary Information
Summary information should include:
· Summary table of validation reports, including analytical method validation, partial
revalidation, and cross-validation reports. The table should be in chronological sequence,
and include assay method identification code, type of assay, and the reason for the new
method or additional validation (e.g., to lower the limit of quantitation).
· Summary table with a list, by protocol, of assay methods used. The protocol number,
protocol title, assay type, assay method identification code, and bioanalytic report code
should be provided.
· A summary table allowing cross-referencing of multiple identification codes should be
provided (e.g., when an assay has different codes for the assay method, validation reports,
17
and bioanalytical reports, especially when the sponsor and a contract laboratory assign
different codes).
B. Documentation for Method Establishment
Documentation for method development and establishment should include:
· An operational description of the analytical method
· Evidence of purity and identity of drug standards, metabolite standards, and internal
standards used in validation experiments
· A description of stability studies and supporting data
· A description of experiments conducted to determine accuracy, precision, recovery,
selectivity, limit of quantification, calibration curve (equations and weighting functions used, if
any), and relevant data obtained from these studies
· Documentation of intra- and inter-assay precision and accuracy
· In NDA submissions, information about cross-validation study data, if applicable
· Legible annotated chromatograms or mass spectrograms, if applicable
· Any deviations from SOPs, protocols, or GLPs (if applicable), and justifications for
deviations
C. Application to Routine Drug Analysis
Documentation of the application of validated bioanalytical methods to routine drug analysis
should include:
· Evidence of purity and identity of drug standards, metabolite standards, and internal
standards used during routine analyses
· Summary tables containing information on sample processing and storage. Tables should
include sample identification, collection dates, storage prior to shipment, information on
shipment batch, and storage prior to analysis. Information should include dates, times,
sample condition, and any deviation from protocols.
· Summary tables of analytical runs of clinical or preclinical samples. Information should
include assay run identification, date and time of analysis, assay method, analysts, start and
18
stop times, duration, significant equipment and material changes, and any potential issues or
deviation from the established method.
· Equations used for back-calculation of results
· Tables of calibration curve data used in analyzing samples and calibration curve summary
data
· Summary information on intra- and inter-assay values of QC samples and data on intra- and
inter-assay accuracy and precision from calibration curves and QC samples used for
accepting the analytical run. QC graphs and trend analyses in addition to raw data and
summary statistics are encouraged.
· Data tables from analytical runs of clinical or preclinical samples. Tables should include
assay run identification, sample identification, raw data and back-calculated results,
integration codes, and/or other reporting codes.
· Complete serial chromatograms from 5B20% of subjects, with standards and QC samples
from those analytical runs. For pivotal bioequivalence studies for marketing,
chromatograms from 20% of serially selected subjects should be included. In other studies,
chromatograms from 5% of randomly selected subjects in each study should be included.
Subjects whose chromatograms are to be submitted should be defined prior to the analysis
of any clinical samples.
· Reasons for missing samples
· Documentation for repeat analyses. Documentation should include the initial and repeat
analysis results, the reported result, assay run identification, the reason for the repeat
analysis, the requestor of the repeat analysis, and the manager authorizing reanalysis.
Repeat analysis of a clinical or preclinical sample should be performed only under a
predefined SOP.
· Documentation for reintegrated data. Documentation should include the initial and repeat
integration results, the method used for reintegration, the reported result, assay run
identification, the reason for the reintegration, the requestor of the reintegration, and the
manager authorizing reintegration. Reintegration of a clinical or preclinical sample should be
performed only under a predefined SOP.
· Deviations from the analysis protocol or SOP, with reasons and justifications for the
deviations
19
D. Other Information
Other information applicable to both method development and establishment and/or to routine
sample analysis could include:
· Lists of abbreviations and any additional codes used, including sample condition codes,
integration codes, and reporting codes
· Reference lists and legible copies of any references
· SOPs or protocols covering the following areas:
· Calibration standard acceptance or rejection criteria
· Calibration curve acceptance or rejection criteria
· Quality control sample and assay run acceptance or rejection criteria
· Acceptance criteria for reported values when all unknown samples are assayed in
duplicate
· Sample code designations, including clinical or preclinical sample codes and
bioassay sample code
· Assignment of clinical or preclinical samples to assay batches
· Sample collection, processing, and storage
· Repeat analyses of samples
· Reintegration of samples
20
GLOSSARY
Accuracy: The degree of closeness of the determined value to the nominal or known true value under
prescribed conditions. This is sometimes termed trueness.
Analyte: A specific chemical moiety being measured, which can be intact drug, biomolecule or its
derivative, metabolite, and/or degradation product in a biologic matrix.
Analytical run (or batch): A complete set of analytical and study samples with appropriate number of
standards and QCs for their validation. Several runs (or batches) may be completed in one day, or one
run (or batch) may take several days to complete.
Biological matrix: A discrete material of biological origin that can be sampled and processed in a
reproducible manner. Examples are blood, serum, plasma, urine, feces, saliva, sputum, and various
discrete tissues.
Calibration standard: A biological matrix to which a known amount of analyte has been added or
spiked. Calibration standards are used to construct calibration curves from which the concentrations of
analytes in QCs and in unknown study samples are determined.
Internal standard: Test compound(s) (e.g. structurally similar analog, stable labeled compound)
added to both calibration standards and samples at known and constant concentration to facilitate
quantification of the target analyte(s).
Limit of detection (LOD): The lowest concentration of an analyte that the bioanalytical procedure
can reliably differentiate from background noise.
Lower limit of quantification (LLOQ): The lowest amount of an analyte in a sample that can be
quantitatively determined with suitable precision and accuracy.
Matrix effect: The direct or indirect alteration or interference in response due to the presence of
unintended analytes (for analysis) or other interfering substances in the sample.
Method: A comprehensive description of all procedures used in sample analysis.
Precision: The closeness of agreement (degree of scatter) between a series of measurements
obtained from multiple sampling of the same homogenous sample under the prescribed conditions.
Processed: The final extract (prior to instrumental analysis) of a sample that has been subjected to
various manipulations (e.g., extraction, dilution, concentration).
21
Quantification range: The range of concentration, including ULOQ and LLOQ, that can be reliably
and reproducibly quantified with accuracy and precision through the use of a concentration-response
relationship.
Recovery: The extraction efficiency of an analytical process, reported as a percentage of the known
amount of an analyte carried through the sample extraction and processing steps of the method.
Reproducibility: The precision between two laboratories. It also represents precision of the method
under the same operating conditions over a short period of time.
Sample: A generic term encompassing controls, blanks, unknowns, and processed samples, as
described below:
Blank: A sample of a biological matrix to which no analytes have been added that is used to
assess the specificity of the bioanalytical method.
Quality control sample (QC): A spiked sample used to monitor the performance of a
bioanalytical method and to assess the integrity and validity of the results of the unknown
samples analyzed in an individual batch.
Unknown: A biological sample that is the subject of the analysis.
Selectivity: The ability of the bioanalytical method to measure and differentiate the analytes in the
presence of components that may be expected to be present. These could include metabolites,
impurities, degradants, or matrix components.
Stability: The chemical stability of an analyte in a given matrix under specific conditions for given time
intervals.
Standard curve: The relationship between the experimental response value and the analytical
concentration (also called a calibration curve).
System suitability: Determination of instrument performance (e.g., sensitivity and chromatographic
retention) by analysis of a reference standard prior to running the analytical batch.
Upper limit of quantification (ULOQ): The highest amount of an analyte in a sample that can be
quantitatively determined with precision and accuracy.
Validation:
Full validation: Establishment of all validation parameters to apply to sample analysis for the
bioanalytical method for each analyte.
22
Partial validation: Modification of validated bioanalytical methods that do not necessarily call
for full revalidation.
Cross-validation: Comparison validation parameters of two bioanalytical methods.