1.0 PURPOSE :
This SOP provides an authorized procedure for cleaning, calibration and validation of BOD incubator.
2.0 SCOPE :
This SOP covers the cleaning, calibration and validation method for BOD incubator (QCI – 43).
3.0 RESPONSIBILITY :
Microbiologist.
4.0 ROCEDURE
4.1.0 CALIBRATION :
Calibration of temperature indicator view controller to be carried out once in a year by external agency or whenever any maintenance done on the apparatus.
4.1.1 :Maintain the record of calibration.
4.2.0 VALIDATION
4.2.1 :Perform the validation of B.O.D. Incubator once in a year.
4.2.2 :Follow the operating procedure for operation after altering set temperature perform the temperature mapping inside the chamber.
4.2.3 :Five calibrated thermometers are kept in different places inside the incubator.
4.2.4 :All the thermometers should maintain the temperature of
± 2 °C of set temperature for 8 hours.
4.2.5 :Maintain the records of validation.
4.3.0 CLEANING METHOD
4.3.1 Disconnect the BOD Incubator from main electrical supply.
4.3.2 Clean the outside and inside of incubator with dry non shredding cloth. Wipe with IPA 70% v/v.
4.3.3 Maintain the log book for cleaning.
4.3.4 Frequency : Once in a week.
validation refers to establishing documented evidence that a process or system, when operated within established parameters, can perform effectively and reproducibly to produce a medicinal product meeting its pre-determined specifications and quality attributes
Subscribe to:
Post Comments (Atom)
Pharmaceutical Validation Documentation Requirements
Pharmaceutical validation is a critical process that ensures that pharmaceutical products meet the desired quality standards and are safe fo...
-
K. Dashora, D. Singh, Swarnlata Saraf and S. Saraf *. Institute of Pharmacy, Pt.RavishankarShuklaUniversity, Raipur 492 010. *Author for ...
-
Validation of the Autoclave is classified into the following 1.0 OQ – Operational Qualification 2.0 PQ – Performance Qualification The valid...
-
Cold storage is a relatively simple cold room that is commonly used to store material between 2[degrees] to 8[degrees]C. Such cold rooms a...
11 comments:
nice blog.MACK PHARMATECH PVT. LTD. PROVIDES Manufacturers of Pharma And Lab Equipments of BOD Incubator .
SGM BOD incubators are designed to meet customized requirements of different customers, SGM as an experienced BOD incubator manufacturer in India, produces standard as well as customized BOD incubator models. These BOD Incubators come in various sizes and capacities such from 36 liters to 1000 liters. They are available in different designs for excellent quality performance.
For more please visit:http://www.lab360.co.in/bod-incubator.htm
Thank you!!!!!!!!!!!!!
Great Information about Calibration of B.O.D Incubator.
We are also manufacturer and Supplier of BOD Incubator and other medical euqipment like Autoclave, Stability chambers, ovens, Laminar Air Flow and many more. Visit our website for more information. - http://www.tancolabproducts.com.
very nice!
You have shared a great information. It is a very useful blog for the users of BOD.
Must read this blog Guys.
For Autoclave Supplies , visit - Autoclave Supplier
Very informative and unique tips dear. Thanks for sharing.
Click here for more information's bod incubator manufacturers
Good concept
Good concept
Very Nice Post. Happy to introduce our services below
Modular Operation Theatre Manufacturers in Chennai
Clean Room Manufacturers in Chennai
Pass Box Manufacturers in Chennai
Air Shower Manufacturers in Chennai
Fume Hood Manufacturers in Chennai
Good and very useful info
Nice post
Storage Tank Manufacturers Chennai
Reactors Manufacturers in Chennai
Pressure vessels Manufacturers in Chennai
Heat Exchangers Manufacturers in Chennai
Piping fabricators Manufacturers in Chennai
Post a Comment